LSP01-1Y Single Channel Syringe Pump
Insurance: 5 x 20 mm, 250 V ~, FAST, 1A
Driver device: Micro -processor control 16 Subsidity step into the motor, drive the screw connecting to the synchronous belt
Polo the distance between each micro -step (1/16 Step): 0.156 Micron
Each micro -step injection liquid (1/16 Step): 0.0919µL (60ml BD Syringe)
Maximum step: 867 Steps/SEC
Minimum step: 1Step/30 SECS
Fast speed range: min: 5um/min max: 13 cm/min
Flow range: 0.831 nl/min to 150.5 ml/min (60 ml syringe)
Linear thrust: 280 cows
Dimensions: 280*325*158.5 (mm)
Weight: 8.2kg
Temperature: 5 ° C -40 ° C (41 ° F -104 ° F)
Humidity: 20% -80% RH
Running mode: independence, synchronization and constant current mode. In the independence and synchronization mode, there are five working methods: irrigation, extraction, first pouring, then drawing, first extraction, then irrigation, and continuous.
main feature
▲ Single -channel irrigation/extraction injection pump is mainly used in the biological laboratory and integrated structure. (Can install a syringe with different specifications of the card, the proofer, suitable for high -precision, small flow liquid transmission.)
▲ Inner diameter input function: you can select the syringe or directly enter the inner class of the syringe from the list
▲ The human -machine interface is friendly, large -screen LCD display, digital knobs are used in combination with film keys, easy to operate and fast
▲ Has the power loss memory function: 1.EEPROM save the setting parameters. After re -powering, no need to be reset 2. Running in the flow mode.
▲ traffic jam protection function: When the advancement of the injection pump is blocked during the work process, the injection pump will stop the work of the promotion agency issuing a whistle alarm
▲ Using the RS485 communication bus can be connected to the upper machine and control it through the background software
▲ External control interface: with input/output control function
▲ Line volume calibration function: more accurate traffic can be obtained through the calibration procedure
▲ Make the syringe protection function: By adjusting the position of the block, it can prevent the syringe from being damaged
Flow parameter
The inner diameter of 10 μl of the syringe is 0.5mm, and the reference flow range is 0.001-0.0255 μL/min-ml/min
The inner diameter of 25 μl of the syringe is 0.8mm, and the reference flow range is 0.0025-0.0653 μL/min-ml/min
The inner diameter of 50 μl of the syringe is 1.1mm, and the reference flow range is 0.0048-0.1235 μL/min-ml/min
The inner diameter of 100 μl of the syringe is 1.6mm, and the reference flow range is 0.0101-0.2614 μL/min-ml/min