Emailrongbaipump@foxmail.com
LSP01-1Y Single Channel Syringe Pump
1 Channel
This product cannot be used in human clinical trials, please be informed!
single channel
304 stainless steel appearance;
128×64 LCD display parameters and various working states;
Fully digital membrane buttons and rotary coding switch control;
Insurance 5x20mm,250V~,Fast,1A;
stepper motor;
Linear thrust ≤280N;
Operation mode: perfusion, extraction, perfusion first and then extraction, extraction first and then perfusion, continuous
The RS485 communication bus can be used to connect to the host computer;
    Features & Benefits
    Brochure
    Video
    Related Products
Technical Specifications
Number of Channels 1
Working Mode Infusion / Withdraw ( Infusion, withdrawal, infusion / withdrawal, withdrawal / infusion, continuous )
Syringe Size 10 uL to 60 mL
Syringe Selection The syringe can be selected in the manufacturer table which includes manufacturer, material and size or input the inner diameter of the syringe barrel directly
Display 128x64 graphic LCD
Display Setting Volume, flow rate, linear speed
Setting Mode Rotary coded switch and membrane keypad
Fast forward / backward Infuse or withdraw liquid at full speed
Advance per Micro-step 0.156 micron
Infusion Volumn per Microstep 0.0919µl (60 mL BD Syringe)
Max. Step Rate 867 steps/sec
Min. Step Rate 1 step/30 secs
Max. Linear Rate 130 mm/min
Min. Linear Rate 5 µm/min
Linear Force ≤280N
Accuracy ≤ 0.5% error in the condition of ≥ 30% of max. infusion distance
Power Supply AC 90-265V  50Hz/60Hz
Power consumption < 50W
Operating temperature 5°C - 40°C (41°F - 104°F)
Relative humidity 20% - 80% RH
Dimensions 280mm*218mm*138.5mm
Weights 4.5 Kg
Communication Function RS485
External Control Function Input / output control
Power-off Memory Storing the running parameters automatically
Calibration Function Acquire accurate volume through calibration
Protection Function The pump will stall and give an alarm when the drive structure of the pump is blocked
Syringe Protection Adjust syringe rest to prevent syringe from damaging
LSP01-1Y Single Channel Syringe Pump